Barel et al.                                                                                                                                                                                          Ultrasound body contouring

           variability, it terms of skin quality and elasticity,   the power output and the magnitude of the effect.
           can  have  an  effect. In  addition,  circumference   An interesting finding was a minimal increase of
           measurement is a rather crude measurement. None of   thigh circumference in some groups. It is unclear if
           the patients underwent prior liposuction, and therefore   beginning a process of body contouring encourages
           no comparison can be made with invasive body       patients to change their life-style to achieve better
           contouring of any kind. Another important factor is that   overall results, or if it permits them to continue their
           patient’s lifestyle effect was not a part of this study;   current habits while relying on technology to correct
           the only control was weight and thigh circumference.   excessive food intake and a lack of physical activity.
           No strict protocol for diet and physical activity was   The low number of adverse events in our cohort is
           taken into consideration. A final item for consideration   consistent with the high safety profile of the device
           is that study-related patient satisfaction in aesthetic   reported  in  extensive  pre-clinical [10]   and  clinical
           procedures may not be comparable to real-life data   studies. [11-14]
           from patients who have actually paid for the procedure.
                                                              Subjective feelings of the patients play a critical role
           The findings indicate that both power levels, with 3   in the outcome of aesthetic procedures, aside from
           treatment sessions, led to a statistically significant   objective evidence of the safety and effectiveness.
           reduction in circumference of the treated area     Therefore, subjects completed a questionnaire
           relative to baseline. Reductions were noted after   evaluating their satisfaction with the results as well as
           each treatment session and overall, after all sessions   the perception of others. Less than half the subjects in
           were completed. The high-power setting was more    both groups noted a favorable visible change following
           effective than the low-power setting in reducing the   the first treatment. However, on later follow-up visits,
           circumference of the treated area after the 1st session   those rates nearly doubled.
           and after the 3rd session.
                                                              In summary, multiple treatments over time with the
           As mentioned, both study groups showed a statistically   Contour I™ device lead to a cumulative reduction in
           significant reduction in circumference of the treated   abdominal circumference. The effect is greater when
           area compared to baseline at all time points, with the   high acoustic power is used. Patient satisfaction was
           exception of the low-power group on day 14 (P = 0.113).   found to be related to the measured circumference
           Our assumption is that this data results from tissue   reduction. The findings support the use of the
           edema.                                             noninvasive Contour I™ as a safe, painless non-
                                                              invasive means for body contouring with proven efficacy
           The decision to use circumference measurement only,   in circumference reduction, correlated to the power of
           and not imaging studies, was based on two reasons.   ultrasonic energy delivered. Due to the limitation of the
           First, subjects normally measure their circumference   study, more studies, comparing different modalities in
           by wearing trousers or skirts, and the aim of the   non-invasive body contouring, are needed. Objective
           treatment is to produce erect posture contouring.   non-positional related measurements in form of
           Secondly, the supine posture in magnetic resonance   accurate three-dimensional photography with volume
           imaging or ultrasound scan does not correlate exactly   calculation may be of future assistance.
           with circumference measurement in the erect position
           since fat distributes differently in the supine position.   Financial support and sponsorship
                                            ®
           Later studies done by Ultrashape  have failed to   The study was supported by Ultrashape , Syneron .
                                                                                                   ®
                                                                                                            ®
           show a significant correlation between these two   The study sponsors participated in study design and
           measurements.                                      analysis of data.
           To isolate the net effect of the device on the specific   Conflicts of interest
           area treated and to eliminate the possibility of bias   There are no conflicts of interest.
           in the interpretation of the results, the circumference
           of the untreated thighs from each subject was used   Patient consent
           as an internal control. The results showed that
           the reduction in circumference of the treated area   Informed consent was obtained from participants.
           relative to the untreated area (true net effect) was
           statistically significant at all-time points. This finding   Ethics approval
           was supported by the cumulative effect of the multiple   The study was approved by the Helsinki Committee
           treatments on the reduction in circumference of    of Rabin Medical Center and Sackler Faculty of
           the treated area as well as the correlation between   Medicine, Tel Aviv University.