Page 133 - Read Online
P. 133

Foppiani et al. Plast Aesthet Res 2023;10:53  https://dx.doi.org/10.20517/2347-9264.2022.137                                Page 7 of 14

               Table 2. Monitoring protocols in thermography and oximetry groups
                Lead author and  Diagnostic  Diagnostic tool  Time frame of  Protocol for diagnostic tool  Cut-off values used for concern  Length of
                publication date  method                use                                                                           monitoring period
                        [19]
                Thiessen et al  .   Thermography  Unspecified  Preoperative,   Once preoperative to determine perforators.   N/a      24-48 h
                2020                                    intraoperative and  Intraoperatively, first after perforator dissection to confirm
                                                        postoperative  patency, then a second after the microvascular
                                                                     anastomosis, then a third after flap inset
                                                                     Postoperatively, 2 measurements taken 1-2 days following
                                                                     surgery
                Saxena et al. [20]    Thermography  FLIR A320 IR thermal   Postoperative  Measurement immediately after the procedure and one   N/a  24 h
                2019                    camera                       day (24 h) after the procedure
                Phillips et al. [21]    Thermography  FLIR one device   Intraoperative,   Intraoperatively after isolation on its vascular pedicle, at   N/a  N/a
                2020                    connected to an iPhone 7  postoperative  max ischemia before anastomosis, in 5-minute intervals
                                        smartphone                   after completion of microvascular anastomosis, and before
                                                                     leaving the OR Postoperatively, used whenever there was
                                                                     concern for flap viability
                        [22]
                Lindelauf et al.     Oximetry  Foresight MC-2030   Preoperative and   Preoperative baseline measurements were performed. A   N/a  24 h
                2021                    oximeter        postoperative  new sensor was positioned postoperatively on the
                                                                     transplanted tissue. In unilateral procedures, postoperative
                                                                     StO2 values of the native breast were also obtained.
                                                                     Measurements were continued for 24 h
                        [23]
                Johnson et al.     Oximetry  ViOptix    Intraoperative and  Intraoperatively following anastomosis. Postoperatively,   Any change 10% or greater  Though discharge
                2021                                    postoperative  hourly checks by nursing staff until the second                with a mean of 4.8
                                                                     postoperative morning, followed by every other hour              days
                                                                     monitoring for the second to third postoperative days, and
                                                                     every fourth-hour monitoring from the third postoperative
                                                                     morning through discharge
                        [24]
                Pelletier et al.     Oximetry  ViOptix   Postoperative  Measurements every 4-6 h until discharge  An StO2 level below 30% or a drop in   Until discharge with a
                2011                                                                                       StO2 level of > 20% per hour lasting for   mean of 3.1 days (ICU
                                                                                                           30 minutes                 group) and 2.7 days
                                                                                                                                      (floor group)
                      [25]
                Ricci et al.  Oximetry  ViOptix         Postoperative  Monitored continuously with tissue oximetry for three   A rapid 20-point drop from baseline in 1 h  72 h
                2017                                                 consecutive days, beginning immediately following the   or an absolute recording < 30 percent
                                                                     procedure
                       [26]
               Ozturk et al.     Oximetry  ViOptix      Intraoperative and  Readings were recorded prior to extubation, after   N/a   36 h
               2014                                     postoperative  extubation and every 4 h for the next 36 h
                      [27]
                Saad et al.  Oximetry   T-Stat tissue oximeter by  Postoperative  Tissue oximetry readings were recorded immediately at   N/a  24 h
                2020                    Spectros                     the completion of the reconstruction at hours 1, 2, 3, 4, 6,
                                                                     12, and 24
                Salgarello et al. [28]    Oximetry  Somanetics INVOS   Postoperative  Measurements recorded continuously for 48 h starting in   An rSO2 value of 30% or drop rate in   48 h
                2018                    5100C Cerebral/Somatic       the post-anesthesia care unit         rSO2 by 20%
                                        Oximeter (Covidien)
                         [29]
                Carruthers et al.    Oximetry  ViOptix   Intraoperative and  Probe applied intraoperatively to skin paddle and remained  N/a  Until discharge with a
                2019                                    postoperative  on until discharge, measurements recorded continuously         mean of 3.4 days
   128   129   130   131   132   133   134   135   136   137   138